7
The complimentary colour method of selecting wavelength may not be applicable in all situations, perhaps
because the solution has no distinct colour, or the solution is of a complex nature and the absorbing species
of interest is not the predominant colour. In these circumstances it will be necessary to scan across the
spectrum to determine points of maximum absorbance.
Situations also exist where solutions will absorb at more than one wavelength. If this situation is encoun-
tered it is usually best to select the wavelength which gives maximum absorbance.
It is always good practice to verify linearity (i.e; concentration vs reading) of the methodology at the se-
lected wavelength. Where more than one absorbing wavelength is available, it is probable that one will
offer better linearity characteristics than others.
To perform measurements the following procedures should be carried out:
Transmission Mode
This mode is selected by moving the cursor to the %T menu using the LEFT or RIGHT arrow keys. The
primary display will show the transmittance, with %T units. The wavelength can be adjusted using the UP
and DOWN arrow keys. A calibration routine is initiated by pressing the CAL key. The routine performs a
zero% transmission calibration followed by a 100% transmission calibration. A blank (clear) solution
should be used for this calibration procedure. (An internal shutter is automatically activated to perform the
zero% setting and this part of the routine is therefore independent of the solution in the light path).
The following error codes are possible after a calibration:
Err 1 Dark cal error. This error occurs when the 6300 closes the shutter to block light entering the sample
chamber. If the detector output does not fall to a level normally associated with a dark cal then this error is
indicated. The most likely cause is that the sample chamber lid is not closed.
Err 2 Light cal error. This error indicates that there is insufficient light to calibrate to 100%. The most likely
cause is that light at the selected wavelength is being absorbed by a sample in the sample chamber. This
error can also be caused if the lamp has failed.
A calibration resets the sample number to unity.
Absorbance Mode
This mode is selected by moving the cursor to the ABS mode using the LEFT or RIGHT arrow keys. The
primary display will show the absorbance, with ABS units. The wavelength can be adjusted using the UP
and DOWN arrow keys. A calibration routine is initiated by pressing the CAL key.The routine performs a
zero% transmission calibration followed by a 0.000 Absorbance calibration. A blank (clear) solution
should be used for this calibration procedure. (An internal shutter is automatically activated to perform the
zero% setting and this part of the routine is therefore independent of the solution in the light path). The
instrument will display overrange
(1. ABS) absorbance during the dark calibration. The possible error conditions are the same as for the
transmission mode.
6300/REV B/12-96
Vista de pagina 9
1 2 ... 5 6 7 8 9 10 11 12 13 14 15 16 17

Comentarios a estos manuales

Sin comentarios

Precor TRM 800 manuals

Owner’s manuals and user’s guides for Sports and recreation Precor TRM 800.
We providing 1 pdf manuals Precor TRM 800 for download free by document types: User Manual






More products and manuals for Sports and recreation Precor

Models Document Type
RECUMBENT RBK 800 User Manual   Precor RECUMBENT RBK 800 User Manual, 100 pages
Commercial Cycles C846i User Manual   Precor Commercial Cycles C846i User Manual, 68 pages
C932 User Manual   Precor C932 User Manual, 60 pages
S3.23 User Manual   Precor S3.23 User Manual, 24 pages
Ellipticals EFX532i User Manual   Precor Ellipticals EFX532i User Manual, 96 pages
946l User Manual   Precor 946l User Manual, 101 pages
EFX5.21s User Manual   Precor EFX5.21s User Manual, 31 pages
EFX546i User Manual   Precor EFX546i User Manual, 114 pages
UBK 800-SERIES User Manual   Precor UBK 800-SERIES User Manual, 100 pages
ELLIPTICAL FITNESS CROSSTRAINERS EFX 835 User Manual   Precor ELLIPTICAL FITNESS CROSSTRAINERS EFX 835 User Manual, 104 pages
UBK 815 User Manual   Precor UBK 815 User Manual, 92 pages
LOW-IMPACT TREADMILLS 9.33 User Manual   Precor LOW-IMPACT TREADMILLS 9.33 User Manual, 86 pages
UBK 800 User Manual   Precor UBK 800 User Manual, 52 pages
EFX 5.21i User Manual   Precor EFX 5.21i User Manual, 58 pages
EFX 883 User Manual   Precor EFX 883 User Manual, 104 pages
776i User Manual   Precor 776i User Manual, 112 pages
Commercial Stretch Trainer C240 User Manual   Precor Commercial Stretch Trainer C240 User Manual, 16 pages
EFX 556i User Manual   Precor EFX 556i User Manual, 70 pages
C934 User Manual   Precor C934 User Manual, 60 pages
EFX 5.25 User Manual   Precor EFX 5.25 User Manual, 64 pages