Brookfield Engineering Labs., Inc. Page 60 M/02-212
BREAK: SPEED MAX
Cause: Maximum speed exceeded during carrying out a measurement.
a) You have selected too high shear stress or too high torque value for the measuring
substance.
b) You have started torque- or shear stress measurement without substance in measuring
system.
Message:
What to do:
Push key.
Select smaller shear stress or torque values for this measurement.
BREAK: Speed max!